Need Help? Chat with us on WhatsApp - We're here to assist you!
FAST DELIVERY FROM GERMANY TO WORLDWIDE
biolab-q6000-fdm-series-fuel-dilution-meter_459328.jpg
biolab-q6000-fdm-series-fuel-dilution-meter_459328.jpg

Quantity :

Request an Offer

97% customer satisfaction based on last 45 days of sales

Biolab Q6000 FDM Series Fuel Dilution Meter

Ask for Stock

Brand:
Product Code:
Q6000 FDM Series

Order Processing & Delivery Information:
As we supply industrial equipment, product availability and delivery timelines may vary. Once your order is placed, our sales team will contact you within 24 business hours (excluding weekends and holidays) to confirm stock status and estimated delivery time.



🎯 Need a Quote for Biolab Q6000 FDM Series Fuel Dilution Meter?

Get the best price and fastest delivery for Biolab Q6000 FDM Series Fuel Dilution Meter from Biolab. Imtek Engineering offers worldwide industrial equipment supply with expert support.


Details:

Fuel dilution in lubrication oil can cause serious engine damage.  In diesel engine crankcases, for example, contamination can be caused by excessive idling, defective injectors and loose connections.  Fuel dilution can also cause intercooler leakage in aircraft hydraulic systems.
The Q6000 Fuel Dilution Meter (FDM) is a portable fuel dilution meter that can be used in the laboratory or in the field to provide rapid and accurate measurements of fuel contamination in engine oil.
The Q6000 was designed with a patent-protected sampling method using an innovative “fang” design.  The sampling occurs using a headspace system with repeatability of <+5 percent RSD up to 15 percent fuel content.  The headspace vapor sampling approach also eliminates the need to heat fuel-laden samples and does not require the use of solvents.
Developed in collaboration with the U.S. Navy, the Q6000 detects contamination in less than two minutes, allowing operators to make immediate decisions. The device requires just a small sample size in a disposable vial.  The analysis results are displayed in percent fuel so operators can make rapid decisions about maintenance actions.  In addition, the intuitive touchscreen and audio prompts guide the operator through the testing process.



Features & Benefits


Conforms to ASTM D8004 – “Standard Test Method for Fuel Dilution of In-Service Lubricants Using Surface Acoustic Wave Sensing”
Innovative headspace sampling fang design ensures excellent repeatability (<= 5% RSD).
0.2 to 15% fuel dilution measurement range.
Disposable FDM vials eliminate carryover contamination.
Onboard memory allows use of up to three stored calibration profiles.
Sample ID entry and data file export via USB cable.
Battery powered, lightweight, optional hard case for transit.



Fast and easy to use


Small sample required in disposable FDM vial – no solvents needed for cleaning.
Test results within a minute for direct measurement of fuel dilution.
Result in % fuel displayed, allowing immediate maintenance actions.
Color touchscreen.
Audio prompt feature guides operators through testing.

Fuel Dilution Meter